 ==== Error 11 - Relay Test Fault - Installation error or possibly relay failure ==== ==== Error 11 - Relay Test Fault - Installation error or possibly relay failure ====
  
-Error 11 indicates a failed automatic self-test of the MultiPlus-II or EasySolar-II relays. It is often a result of an installation or wiring issue, though it can also be triggered by a hardware failure.+Error 11 indicates a failed automatic self-test of the MultiPlus-II or EasySolar-II relays. It is often a result of an installation or wiring issue, though it can also be triggered by a hardware failure ​(eg inoperable relays).
  
 If a grid code is selected in VEConfigure,​ then the backfeed relays are tested every time before connecting to the mains. If a grid code is selected in VEConfigure,​ then the backfeed relays are tested every time before connecting to the mains.
  
-The other tests that raise Error 11 when they fail concern the Ground Relay. During inverter operation, disconnected from mains, the relay is continuously tested to be closed. And it is tested to open properly during switchover to mains. The ground relay tests are always performed: with or without a configured grid code. Unless the relay is disabled in VEConfigure. See product manual for more information regarding the ground relay.+Ground relays are also tested and can trigger an Error 11. During inverter operation, disconnected from mains, the relay is continuously tested to be closed. And it is tested to open properly during switchover to mains. The ground relay tests are always performed: with or without a configured grid code. Unless the relay is disabled in VEConfigure. See product manual for more information regarding the ground relay.
  
 === Steps to take: === === Steps to take: ===
  
-Step 1: Make sure that the units have been [[updating_firmware:​updating_ve.bus_products|updated to the latest firmware version]]. There have been various VE.Bus error 11 related improvements in the firmware.+Step 1: Make sure that the units have been [[updating_firmware:​updating_ve.bus_products|updated to the latest firmware version]]. There have been various VE.Bus error 11 related improvements in the firmware. This may correct the error, and if it does not, it will provide additional information to assist troubleshooting.
  
-Step 2: Verify that system is properly wired to your installation and your local codes. Remember to check the earth connection on the unit AC input. Earth should be grounded according to your local grid rules. Do not connect Earth to Neutral ​on the unit. This will create a safety hazard and will prevent RCD devices from working. Contact ​licensed electrician if in doubt. ​+Step 2: Verify that system is properly wired to your installation and your local codes. Remember to check the earth connection on the unit AC input. Earth should be grounded according to your local grid rules. Do not connect Earth to Neutral ​in the unit itself. This will create a safety hazard and may prevent RCD devices from working. Contact ​another ​licensed electrician if in doubt. ​
  
-Step 3: Disconnect the wiring from the ACout terminals on each inverter/​charger. Which, in case of a parallel or three phase system, isolates the ACout connections for each unit. If the system then starts normally there might be a problem with a connected load (e.g. a leakage between load neutral and ground).+Step 3: Disconnect the L and N wiring from the ACout terminals on each inverter/​charger. Which, in case of a parallel or three phase system, isolates the ACout connections for each unit. If the system then starts normally there might be a problem with a connected load (e.g. a leakage between loadneutral and ground).
  
-Step 4: Check your Neutral-out connection. ​There should be no external connection ​from Neutral-out to Neutral-in or to Earth. ​If your local grid code requires ​bypass ​between Neutral-in and Neutral-out, ​contact your Victron representative for a solution.+Step 4: Check your Neutral-out connection. ​In most cases there should be no external connection ​between ACin-Neutral ​to ACout-Neutral, ​or to Earth. ​ 
 + 
 +In Australia, it is possible if required to install with connection ​between Neutral-in and Neutral-out ​using AS4777.2:​2015. Select either design configuration in the grid code section while programming the MultiPlus-II with firmware 466 and above. After configurationalways confirm that the system is safe with routine electrical safety checks in both modes (eg test RCD operation with AC input connected and disconnected).
  
 Step 5: Neutral and line input might be inverted. Check the input connections,​ swap them if needed and retry. In this case, “Failing step: 8” will show. Step 5: Neutral and line input might be inverted. Check the input connections,​ swap them if needed and retry. In this case, “Failing step: 8” will show.
  
-=== Unable to solve? ​=== +=== Detailed diagnosis of Error 11 ===
-When the error persists after the all above steps, or when any other Failed step or extra explanation shows, then the unit is most likely faulty. Follow the support procedure.+
  
-Make sure to provide ​photo of the error details shown on the GX Device ​menu.+After updating your VE.Bus inverter/​charger ​to firmware 454 or higher; and running Venus OS v2.23 or later, ​detailed VE.Bus Error 11 report is available in the GX device ​menu, under MultiPlus -> Alarm Status -> VE.Bus Error 11 report.
  
-=== Detailed diagnosis ​of Error 11 === +Examples ​of information ​in the GX Device menus: 
-After updating your VE.Bus inverter/​charger to firmware 454 or higher; and running Venus OS v2.23 or later, a detailed VE.Bus Error report is available ​in the GX device menu, under MultiPlus ​-> Alarm Status -> VE.bus error. The information shown is:+ 
 +{{:​ve.bus:​error11-ccgx3.jpg?​nolink|}} 
 +{{:ve.bus:​error11-ccgx4.jpg?​nolink|}} 
 + 
 + 
 +The information shown is:
  
   * A six digit number; which contains all the details   * A six digit number; which contains all the details
Line 130: Line 136:
  
   * Failing step #8: check that line and neutral have not been swapped. Unlike non-grid code installations;​ line and neutral need to be wired correctly.   * Failing step #8: check that line and neutral have not been swapped. Unlike non-grid code installations;​ line and neutral need to be wired correctly.
-  * PE2 open error or GND Relay Error: check that AC-out Neutral is not wired to Ground. Short circuiting them is not allowed. Also, check that there is no leakage between Neutral and ground, for example because broken consumer. Start by disconnecting all loads: open all the circuit breakers. +  * PE2 closed ​error or GND Relay Error: check that AC-out Neutral is not wired to ground, or AC-in Neutral + ground. Short circuiting them is not allowed ​without specific firmware and grid code. Also, check that there is no leakage between Neutral and ground, for example because broken consumer. Start by disconnecting all loads: open all the circuit breakers. Then test to confirm there is no continuity between ACin-Neutral and ACout-Neutral
-  * AC0/AC1 mismatch and UMains error: ​there has been a fix for that in VE.Bus inverter/​charger firmware version 455.+  * AC0/AC1 mismatch and UMains error: ​this is fixed in VE.Bus inverter/​charger firmware version 455.
  
 Other shown error details, listed below, are not related to any known installation issue and can be caused by a faulty unit or configuration issue. In any case, also when one of below errors shows, to go through all the above steps for any error 11 issue. Other shown error details, listed below, are not related to any known installation issue and can be caused by a faulty unit or configuration issue. In any case, also when one of below errors shows, to go through all the above steps for any error 11 issue.
Line 143: Line 149:
   * PE2 open error.   * PE2 open error.
  
-The report is found by navigating ​to the Device list (top level menu), then going in to the Multi or Quattro inverter/​charger page, then to the bottom, Alarm Status, then to the bottom again, in to the menu VE.Bus Error 11 report. +=== Unable ​to solve? ===
- +
-Now you will see a list of all devices in the system; and their latest VE.Bus Error 11 details. Enter the menu for an explanation of the code. +
- +
-Examples of information in the GX Device menus:+
  
-{{ :ve.bus:error11-ccgx3.jpg?​nolink ​|}} +When the error persists after the all above steps, or when any other Failed step or extra explanation shows, then the unit may be faultyFollow the [[https://www.victronenergy.com/​support|support procedure and contact your dealer.]] Make sure to document full system wiring in a diagram, all steps and tests taken to diagnose and rule out installation error, and provide a photo of the error details shown on the GX Device menu.
-{{ :ve.bus:​error11-ccgx4.jpg?nolink |}}+
 ==== Error 12 - Config mismatch with 2nd mcu ==== ==== Error 12 - Config mismatch with 2nd mcu ====
  
Line 248: Line 249: